Measurements of atmospheric carbonyl sulfide during the NASA Chemical Instrumentation Test and Evaluation Project: Implications for the global COS budget

Johnson, J.E., A.R. Bandy, D.C. Thornton, and T.S. Bates

J. Geophys. Res., 98(D12), 23,443–23,448, doi: 10.1029/92JD01911 (1993)


Atmospheric COS concentrations were measured by three analytical systems during the Chemical Instrumentation Test and Evaluation (CITE 3) project. The three systems all used cryogenic sample preconcentration and gas chromatographic (GC) separation but differed in the method of detection. The FPD system used a flame photometric detector, the MS system used a mass selective detector, and the ECD-S system used a fluorinating catalyst followed by an electron capture detector. With the FPD system, we found a mean COS concentration of 510 ppt over the North Atlantic and 442 ppt over the Tropical Atlantic. With the ECD-S system, we found a mean COS concentration of 489 ppt over the North Atlantic and 419 ppt over the Tropical Atlantic. All three systems registered a latitudinal gradient in atmospheric COS of between 1.6 and 2.0 ppt per degree of latitude, with increasing COS concentrations northward which was similar to the gradient measured by Bingemer et al. (1990). It is difficult to reconcile the measured latitudinal concentration gradient with present theories of the global COS budget since the largest sink of COS is thought to be a flux to land plants, most of which are in the northern hemisphere.